 Embedded Power for Business-Critical Continuity
LPT60-M Series
60 Watts Medical
Total Power: Input Voltage: # of Outputs: 60 - 80 Watts 85 - 264 VAC 120 - 300VDC Triple
Rev. 06.15.07 LPT60-M Series 1 of 3
RoHS
Electrical Specifications
Special Features
* * * * * * * * * * * Medical safety approvals Universal input 3" x 5" footprint Remote sense on main output Built-in EMI filter Low output ripple Adjustable 5V output Overvoltage protection Overload protection 110KHz switching frequency LPX60 enclosure kit available
Input
Input range Frequency Inrush current Input current Efficiency EMI 85-264VAC 120-300VDC 47-440Hz <18A peak @ 115VAC; <36A peak @ 230VAC, cold start @ 25C 1.5A max. (RMS) @ 115VAC 70% typical at full load FCC Class A conducted; CISPR 22 Class A conducted EN55022 Class A conducted; VDE 0878 PT3 Class A conducted <75A max. @ 50/60 Hz, 264VAC input
Safety ground leakage current
Output
Maximum Power Cross regulation Adjustment range Hold-up time Overload protection 0750/EN60601-1 (IEC601) 21310-3336-0015 UL2601 E182560 CSA 22.2 No. 601.1 LR109492 Mark (LVD) Overvoltage protection Remote sense 60W for convection; 80W with 30CFW forced air 2% on output 1; 5% on outputs 2, 3 -5, +10% minimum 20ms @ 60W load, 115VAC nominal line Short circuit protection on all outputs. Case overload protected @ 110% to 145% above peak rating 5.7 to 6.7 VDC on the main output Compensates for 0.5 V lead drop min. Will operate without remote sense connected. Reverse connection protected.
Safety
VDE UL CSA CE
Embedded Power for Business-Critical Continuity
Environmental Specifications
Operating temperature: Electromagnetic susceptibility: Humidity: Vibration: 0 to 50C ambient derate each output at 2.5% per degree from 50 to 70C, -20C startup Designed to meet IEC 801, -2, -3, -4, -5, -6, Level 3 Operating; non-condensing 5% to 95% Three orthogonal axes, sweep at 1 oct/min, 5 min. dwell at four major resonances 0.75G peak 5 Hz to 500 Hz, operational -40 to 85C .04% per C >550,000 hours at full load and 25C ambient conditions
Rev. 06.15.07 LPT60-M Series 2 of 3
Storage temperature: Temperature coefficient: MTBF demonstrated:
Ordering Information
Maximum Load Maximum Load with Convnection with 30CFM cooling Forced Air LPT62-M +5V 0.7A 7A 8A +12V 0.3A 3A 3.5A -12V 0A 0.7A 1A LPT63-M +5V 0.7A 7A 8A +15V 0.3A 2.8A 3.3A -15V 0A 0.7A 1A 1. Peak current lasting <30 seconds with a maximum 10% duty cycle. Model Number Output Voltage Minimum Load Peak Load1 10A 6A 1.5A 10A 4A 1.5A Regulation2 2% 5% 5% 2% 5% 5% Ripple P/P (PARD)3 50mV 120mV 120mV 50mV 150mV 150mV
2. At 25C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings. 3. Peak-to-peak with 20 MHz bandwidth and 10F in parallel with a 0.1F capacitor at rated line voltage and load ranges. 4. Minimum Loads are required.
Embedded Power for Business-Critical Continuity
Mechanical Drawing
Americas
5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698
Rev. 06.15.07 LPT60-M Series 3 of 3
Europe (UK)
Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355
Asia (HK)
16th - 17th Floors, Lu Plaza 2 Wing Yip Street, Kwun Tong Kowloon, Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888
For global contact, visit:
www.astecpower.com www.artesyn.com technicalsupport@astec.com technicalsupport@artesyn.co m
While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions. Printed in US A
Pin Assignments
Connector SK1-1 SK1-3 SK2-1 SK2- 2 SK2- 3 SK2-4 SK2-5 SK2-6 SK201-1 SK201-2 LPT62-M Neutral Line +12V +5V +5V Common Common -12V +Sense -Sense LPT63-M Neutral Line +15V +5V +5V Common Common -15V +Sense -Sense
Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power Embedded Power Inbound Power Integrated Cabinet Solutions Outside Plant Precision Cooling Site Monitoring and Services EmersonNetworkPower. com
Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. (c)2007 Emerson Electric Co.
Mating Connectors
AC Input: DC Outputs: Molex 09-50-8031 (USA) 09-91-0300 (UK) PINS: 08-58-0111 Molex 09-50-8061 (USA) 09-91-0600 (UK) PINS: 08-52-0113
Remote Sense: Molex 22-01-2025 PINS: 08-52-0113 Astec Connector Kit #70-841-006, includes all of the above 1. 2. 3. 4. 5. 6. 7. Specifications subject to change without notice. All dimensions in inches (mm), tolerance is 0.02" (0.5mm) Mounting holes M1 and M2 should be grounded for EMI purposes. Mounting hole M1 is safety ground connection. Specifications are for convection rating at factory settings at 115 VAC input, 25C unless otherwise stated. Warranty: 2 year Weight: 0.75lbs/0.34kg